Company record
Beaver-Visitec International, Inc.
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about Beaver-Visitec International, Inc..
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jan 7 | Class II | Beaver-Visitec International, Inc. Device |
BVI Leos Laser and Endoscopy System. Model Number: OME6000U. Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable). FDA notice Z-1008-2026 |